FDA Adverse Event Malfunction Summary report: N

FLOGARD

MDR report key: 3191212 · Received June 26, 2013

Report

Report Number
1416980-2013-16466
Event Type
Malfunction
Date Received
June 26, 2013
Date of Event
June 1, 2013
Report Date
June 1, 2013
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MX
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS SERVICED AT THE CUSTOMER SITE. THE SAMPLE WAS VISUALLY INSPECTED AND FUNCTIONALLY TESTED. THE REPORTED CONDITION OF AN OCCLUSION ALARM WAS CONFIRMED AS AN F-2 ALARM. THE CAUSE OF THE REPORTED CONDITION WAS DUE TO A DAMAGED LATCH ROLLER. IN ORDER TO RESOLVE THE ISSUE THE LATCH ROLLER WAS REPLACED. THE DEVICE WAS SERVICED AND RESTORED TO A GOOD WORKING CONDITION. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A FOLLOW UP MDR WILL BE SENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT A FLOGARD INFUSION PUMP HAD EXPERIENCED AN OCCLUSION ALARM. THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION WAS REQUESTED AND IS NOT AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
289772 FLOGARD PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1