FDA Adverse Event Malfunction Summary report: N

SPECTRUM INFUSION PUMP

MDR report key: 3191190 · Received June 14, 2013

Report

Report Number
1314492-2013-00791
Event Type
Malfunction
Date Received
June 14, 2013
Date of Event
May 1, 2013
Report Date
May 17, 2013
Manufacturer
BAXTER HEALTHCARE CORP
Product Code
FRN
PMA / PMN Number
K042121
Removal / Correction Number
1314492-080712-002-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED AND EVALUATED BY BAXTER. THE DEVICE EVALUATION CONFIRMED THE FOLLOWING KEYS ARE INOPERABLE: (.), #2, #3, AND #4, #5, #8 AND #9 KEYS CAUSED BY A FAILED KEYPAD. IT WAS OBSERVED THAT WHEN ANY REMAINING FUNCTIONAL KEYS ARE SELECTED, AN AUTOMATIC OUTPUT OF THE #2 KEY WILL OCCUR FOLLOWING THE SELECTED KEY'S FUNCTION (E.G. WHEN THE #1 KEY IS PRESSED 12 WILL BE DISPLAYED. WHEN IN ALPHABETIC MODE AND THE "ABC" KEY IS SELECTED, AD WILL BE DISPLAYED INTERFERING WITH THE MDL DRUG SEARCH). THE FAILED KEYPAD WAS REPLACED. BAXTER HAS INITIATED A CAPA TO INVESTIGATE THE REPORTED EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PUMPS KEYPAD WAS INPUTTING DOUBLE VALUES WHEN A SINGLE KEY IS PRESSED. IT WAS ALSO REPORTED THAT THERE WAS NO PT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
272380 SPECTRUM INFUSION PUMP INFUSION PUMP FRN BAXTER HEALTHCARE CORP 35700

Patients

Seq Age Sex Outcome Treatment
1