FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3191030 · Received June 26, 2013

Report

Report Number
2531779-2013-09042
Event Type
Malfunction
Date Received
June 26, 2013
Report Date
May 30, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. DEVICE EVALUATION: REVIEW OF THE BLACK BOX DATA REVEALED MULTIPLE RECORDS OF ¿POWER ON RESET¿ ON (B)(6) 2013. ON EXAMINATION, THERE WAS EVIDENCE OF MOISTURE INSIDE THE BATTERY COMPARTMENT. DURING INVESTIGATION, POWER LOSS WAS NOT OBSERVED. THE PUMP WAS EXERCISED FOR 24 HOURS WITHOUT ANY ISSUES. A LEAK TEST REVEALED A LEAK AT THE DISPLAY LENS. THE PUMP WAS OPENED FOR INVESTIGATION AND REVEALED EVIDENCE OF MOISTURE CONTAMINATION INSIDE THE PUMP. INVESTIGATION WAS UNABLE TO DUPLICATE THE COMPLAINT.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (INTERMITTENT POWER) ISSUE. THE REPORTER STATED THAT THE PUMP WAS BLANK AT RANDOM. THE REPORTER STATED THAT THE PATIENT HAD A BG OF 200 MG/DL WITH NO SYMPTOMS, WHICH DOES NOT MEET ANIMAS'S CRITERIA FOR AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
289548 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 14 YR