FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 4000 SYSTEM
MDR report key: 3190982
·
Received June 21, 2013
Report
- Report Number
- 3008642652-2013-01649
- Event Type
- Malfunction
- Date Received
- June 21, 2013
- Date of Event
- May 29, 2013
- Report Date
- June 18, 2013
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVAL OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (CHECK THERAPY ELECTRODE ALARMS) WAS CONFIRMED. UPON INVESTIGATION THE BLACK PULSE WIRE WAS OPEN INSIDE OF THE TRUNK CABLE. THE OPEN PULSE WIRE CAUSED THE REPORTED ALARMS. THE ROOT CAUSE FOR THE OPEN PULSE WIRE COULD NOT BE POSITIVELY IDENTIFIED BUT WAS LIKELY EXCESSIVE FORCE ON THE TRUNK CABLE. NO ADVERSE EVENT RESULTED FROM THE OPEN PULSE WIRE. THE PT RECEIVED A REPLACEMENT ELECTRODE BELT.
Description of Event or Problem · 1
A ZOLL PT SERVICE REP CALLED ZOLL CUSTOMER SUPPORT TO REPORT THAT A (B)(6) MALE PT WAS RECEIVING CONSTANT CHECK THERAPY ELECTRODE ALARMS. THE PT WAS ISSUED A REPLACEMENT ELECTRODE BELT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 282325 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 51 YR |