FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 3190982 · Received June 21, 2013

Report

Report Number
3008642652-2013-01649
Event Type
Malfunction
Date Received
June 21, 2013
Date of Event
May 29, 2013
Report Date
June 18, 2013
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVAL OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (CHECK THERAPY ELECTRODE ALARMS) WAS CONFIRMED. UPON INVESTIGATION THE BLACK PULSE WIRE WAS OPEN INSIDE OF THE TRUNK CABLE. THE OPEN PULSE WIRE CAUSED THE REPORTED ALARMS. THE ROOT CAUSE FOR THE OPEN PULSE WIRE COULD NOT BE POSITIVELY IDENTIFIED BUT WAS LIKELY EXCESSIVE FORCE ON THE TRUNK CABLE. NO ADVERSE EVENT RESULTED FROM THE OPEN PULSE WIRE. THE PT RECEIVED A REPLACEMENT ELECTRODE BELT.

Description of Event or Problem · 1

A ZOLL PT SERVICE REP CALLED ZOLL CUSTOMER SUPPORT TO REPORT THAT A (B)(6) MALE PT WAS RECEIVING CONSTANT CHECK THERAPY ELECTRODE ALARMS. THE PT WAS ISSUED A REPLACEMENT ELECTRODE BELT.