ANIMAS INSULIN INFUSION PUMP
Report
- Report Number
- 2531779-2013-09037
- Event Type
- Malfunction
- Date Received
- June 26, 2013
- Report Date
- June 5, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
(B)(6). THE PUMP HAS BEEN RETURNED TO ANIMAS BUT EVALUATION HAS NOT YET BEEN COMPLETED. WHEN EVALUATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME.
FOLLOW-UP # 1 DATE OF SUBMISSION 08/27/2013 - DEVICE EVALUATION:THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS 08/07/2013 WITH THE FOLLOWING FINDINGS:DURING TESTING, THE DISPLAY SCREEN WAS FOUND TO BE BLANK DURING START UP. THE PUMP VIBRATED AND ALARMED WHEN POWERED ON. THE PUMP WAS OPENED AND THE DISPLAY WAS FOUND TO BE CRACKED. A TEST SCREEN WAS INSERTED AND WAS FOUND TO FUNCTION PROPERLY. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
ON (B)(6) 2013, THE DISTRIBUTER CONTACTED ANIMAS, ALLEGING A DISPLAY (DAMAGED) ISSUE. THERE WAS REPORTEDLY INK SPOTS ON THE DISPLAY LENS. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 291032 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |