FDA Adverse Event Malfunction Summary report: N

HOMECHOICE

MDR report key: 3190856 · Received June 26, 2013

Report

Report Number
1416980-2013-16453
Event Type
Malfunction
Date Received
June 26, 2013
Date of Event
May 28, 2013
Report Date
June 1, 2013
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K102936
Removal / Correction Number
1423500-01/08/10-001-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MX
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS EVALUATED AND THE REPORTED ISSUE WAS CONFIRMED THROUGH THE REVIEW OF THE LOGS. THE CAUSE OF THE REPORTED ISSUE WAS DETERMINED TO BE ONE OR MORE CYCLES ADVANCES TO NEXT FILL WHEN SLOW / NO FLOW OCCURRED ABOVE THE MIN DRAIN VOLUME THRESHOLD. IF ADDITIONAL RELEVANT INFORMATION BECOMES AVAILABLE, A FOLLOW UP MEDWATCH WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN EVALUATION OF A RETURNED HOMECHOICE (HC) MACHINE, THERE WAS ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT IDENTIFIED WHICH OCCURRED IN THE THERAPY INITIATED ON (B)(6) 2013 AT 22:05:53. THE HOME PATIENT?S (HP) ULTRAFILTRATION READING OF 1380ML DURING THE NIGHT DRAIN CYCLE FOUR INDICATED THAT THE HP DRAINED 1380ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 2000ML. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
290293 HOMECHOICE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1