FDA Adverse Event Malfunction Summary report: N

PWRD ECH FLEX 60MM

MDR report key: 3190792 · Received June 26, 2013

Report

Report Number
3005075853-2013-03300
Event Type
Malfunction
Date Received
June 26, 2013
Date of Event
May 12, 2013
Report Date
May 13, 2013
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K110385
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION WAS NOT PROVIDED BY THE INITIAL CONTACT. INCORRECT CARTRIDGE SIZE. J5JH6D THE ANALYSIS FOUND THAT ONE PSE60A DEVICE WAS RETURNED IN GOOD VISUAL CONDITION AND WITH THE MANUAL OVERRIDE DOOR OUT OF POSITION; THE OVERRIDE LEVER WAS UP WHICH DENOTES THAT THE KNIFE WAS MANUALLY RETURNED TO HOME POSITION. AFTER THE MANUAL OVERRIDE SYSTEM IS USED THE INSTRUMENT IS DISABLED AND CANNOT BE USED FOR ANY SUBSEQUENCE FIRINGS. AN ECR45M CARTRIDGE RELOAD WAS LOADED IN THE DEVICE; IT WAS NOTED TO BE UNFIRED. IT SHOULD BE NOTED THAT A 60MM DEVICE IS DESIGNED TO WORK ONLY WITH 60MM CARTRIDGES, FOR FURTHER LOADING INSTRUCTIONS PLEASE REFER TO THE ECHELON FLEX 60 POWERED IFU. THE DEVICE WAS DISASSEMBLED TO RESET THE BAILOUT SYSTEM AND THEN, THE INSTRUMENT WAS TESTED FOR FUNCTIONALITY IN THE STRAIGHT POSITION WITH A 60MM TEST CARTRIDGE RELOAD AND ACHIEVED ITS COMPLETE FIRING SEQUENCE WITHOUT ANY DIFFICULTIES. THE STAPLE LINE AND CUT LINE WERE COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORM SHAPE. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROSCOPIC APPENDECTOMY PROCEDURE, ON THE FIRST FIRING THE DEVICE WOULD NOT FIRE. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. NO ADVERSE CONSEQUENCES REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
291603 PWRD ECH FLEX 60MM STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. NA UNK

Patients

Seq Age Sex Outcome Treatment
1 ECR45M