FDA Adverse Event Malfunction Summary report: N

LIGAMAX-5MM ENDO CLIP APPLIER

MDR report key: 3190789 · Received June 26, 2013

Report

Report Number
3005075853-2013-03299
Event Type
Malfunction
Date Received
June 26, 2013
Date of Event
June 19, 2013
Report Date
June 21, 2013
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
FZP
PMA / PMN Number
K050344
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4): INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RETURNED IN GOOD VISUAL CONDITION. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. UPON TESTING, THE DEVICE WAS CYCLED, FED, AND FORMED THE REMAINING CLIPS AS INTENDED DURING ANALYSIS THE JAWS OPEN AND CLOSE AS INTENDED. IN ADDITION THE DEVICE LOCKED OUT AS INTENDED. NO CONCLUSION COULD BE REACHED AS TO WHAT MAY HAVE CAUSED THE REPORTED INCIDENT. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE, JAMMING OCCURRED AND THE JAWS DID NOT OPEN AFTER THE 5TH FIRING. THE DEVICE DID NOT CLAMP THE PATIENT'S TISSUE AT THE TIME. AFTER THAT, THE CLIP DID NOT COME OUT WHEN THE DEVICE WAS FIRED OUT OF THE PATIENT. THE FIRING FORCE WAS HIGHER THAN EXPECTED. THERE WAS NO UNEXPECTED NOISE. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.