ANIMAS VIBE
Report
- Report Number
- 2531779-2013-09019
- Event Type
- Malfunction
- Date Received
- June 26, 2013
- Date of Event
- June 5, 2013
- Report Date
- June 5, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS BEEN RETURNED TO ANIMAS. EVALUATION HAS NOT YET BEEN COMPLETED. WHEN EVALUATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME. (B)(6).
FOLLOW UP #1 SUBMITTED: 08/21/2013. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.DEVICE EVALUATION: ON INVESTIGATION THE DISPLAY SCREEN WAS FULLY FUNCTIONAL AN D FULLY ILLUMINATED WITH NO VISIBLE SIGNS OF DAMAGE, FADING OR DISCOLORATION. INVESTIGATION WAS UNABLE TO DUPLICATE THE COMPLAINT OF A BLANK SCREEN.
THE DISTRIBUTOR CONTACTED ANIMAS ON (B)(6) 2013 AND REPORTED THE DISPLAY SCREEN WAS BLANK/FADED. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 290866 | ANIMAS VIBE | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |