FDA Adverse Event Malfunction Summary report: N

PLV-100

MDR report key: 3190674 · Received June 24, 2013

Report

Report Number
2518422-2013-01274
Event Type
Malfunction
Date Received
June 24, 2013
Date of Event
December 6, 2010
Report Date
December 6, 2010
Manufacturer
RESPIRONICS INC.
Product Code
CBK
PMA / PMN Number
K832467
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE VENTILATOR WAS RETURNED TO THE MFR FOR EVALUATION AND THE CUSTOMER'S COMPLAINT WAS CONFIRMED. THE DEVICE'S BALL SCREW ASSEMBLY AND PISTON CYLINDER WERE REPLACED TO ADDRESS THE ISSUE. DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, AN ISSUE RELATED TO THE POWER BOARD WAS OBSERVED. THE DEVICE'S POWER BOARD WAS REPLACED TO ADDRESS THE ISSUE. THIS REPORT IS BEING SUBMITTED AS PART OF A PROGRAM TO RETROSPECTIVELY REVIEW POSSIBLE DEVICE MALFUNCTIONS THAT DID NOT INVOLVE ANY INJURY, TO DETERMINE WHETHER THEY ARE REPORTABLE UNDER 21 CFR PART 803 AND THE MANUFACTURER'S UPDATED REPORTING PROCEDURES. THIS PROGRAM IS BEING UNDERTAKEN TO ADDRESS FDA WARNING LETTER (B)(4).

Description of Event or Problem · 1

A VENTILATOR WAS RETURNED TO THE MANUFACTURER WITH AN ALLEGATION THAT THE UNIT STOPPED CYCLING. THERE WAS NO HARM OR INJURY REPORTED.