FDA Adverse Event Malfunction Summary report: N

INFINITI VISION SYSTEM

MDR report key: 3190659 · Received June 21, 2013

Report

Report Number
2028159-2013-01202
Event Type
Malfunction
Date Received
June 21, 2013
Date of Event
May 23, 2013
Report Date
May 23, 2013
Manufacturer
ALCON - IRVINE TECHNOLOGY CENTER
Product Code
HQC
PMA / PMN Number
K021566
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE COMPANY REPRESENTATIVE EXAMINED THE SYSTEM AND CONFIRMED THE SYSTEM MESSAGE REPORTED (DETENT FAILURE). THE FOOTSWITCH CABLE WAS REPLACED. THE SYSTEM WAS THEN TESTED AND MET ALL PRODUCT SPECIFICATIONS. THERE WAS NO SAMPLE RETURNED FOR EVALUATION AND NO ADDITIONAL INFORMATION PROVIDED RELATED TO THIS EVENT. FOR THIS REASON, STEPS COULD NOT BE TAKEN TO REPLICATE OR CONFIRM THE REPORTED EVENT. A REVIEW OF COMPLAINTS FOR THE LAST 24 MONTHS DID NOT INDICATE ANY ADDITIONAL SIMILAR REPORTS FOR THIS SYSTEM. THE ROOT CAUSE CANNOT BE DETERMINED CONCLUSIVELY. (B)(4).

Description of Event or Problem · 1

A CUSTOMER REPORTED THAT THERE WAS NO FOOTSWITCH RESPONSE AND A SYSTEM MESSAGE DISPLAYED DURING A PROCEDURE. THE CASE WAS COMPLETED USING A MANUAL TECHNIQUE AND THERE WAS NO IMPACT TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
281912 INFINITI VISION SYSTEM PHACOFRAGMENTATION SYSTEM HQC ALCON - IRVINE TECHNOLOGY CENTER INFINITI NA

Patients

Seq Age Sex Outcome Treatment
1 UNK