FDA Adverse Event Malfunction Summary report: N

RUSCH FIBEROPTIC LARYNGOSCOPE HANDLE

MDR report key: 3190578 · Received June 21, 2013

Report

Report Number
3004365956-2013-00202
Event Type
Malfunction
Date Received
June 21, 2013
Date of Event
May 24, 2013
Report Date
May 30, 2013
Manufacturer
TELEFLEX MEDICAL
Product Code
CCW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

VISUAL INSPECTION: FIVE (5) HANDLES WERE RECEIVED UNDER THE LOT/SERIAL #(B)(4). THE HANDLES DO NOT DISPLAY ANY VISIBLE SIGNS OF DAMAGE OTHER THAN LIGHT SUPERFICIAL SCRATCHES THAT WOULD COME FROM NORMAL WEAR AND TEAR. FUNCTIONAL INSPECTION: PER THE ATTACHED PICTURE IN THIS SAP COMPLAINT, THREE DIFFERENT STYLE BLADES (RUSCH DISPOSABLE GREEN LITE MILL 2, SNAP LIGHT MILL 4, AND EMERALD RUSCH MIL 1.5) WERE ATTACHED TO EACH OF THE FIVE SUSPECTED DEFECTIVE HANDLES AND REMOVED WITHOUT ANY REASONABLE FORCE. ROOT CAUSE: NO FUNCTIONAL ISSUES FOUND.

Description of Event or Problem · 1

THE COMPLAINT WAS REPORTED AS: THE CUSTOMER ALLEGES THAT THEY ARE UNABLE TO REMOVE THE BLADE FROM THE HANDLE WITHOUT EXTRAORDINARY EFFORT. THE HANDLES ARE USED WITH SUNMED, PHILLIPS AND RUSCH BLADES. THE END-USER STATES THAT THE ISSUE OCCURS OVER TIME. NO REPORT OF PT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
283977 RUSCH FIBEROPTIC LARYNGOSCOPE HANDLE FIBEROPTIC LARYNGOSCOPE HANDLE CCW TELEFLEX MEDICAL

Patients

Seq Age Sex Outcome Treatment
1 SUNMED BLADES| RUSCH BLADES| PHILLIPS BLADES