RUSCH FIBEROPTIC LARYNGOSCOPE HANDLE
Report
- Report Number
- 3004365956-2013-00202
- Event Type
- Malfunction
- Date Received
- June 21, 2013
- Date of Event
- May 24, 2013
- Report Date
- May 30, 2013
- Manufacturer
- TELEFLEX MEDICAL
- Product Code
- CCW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
VISUAL INSPECTION: FIVE (5) HANDLES WERE RECEIVED UNDER THE LOT/SERIAL #(B)(4). THE HANDLES DO NOT DISPLAY ANY VISIBLE SIGNS OF DAMAGE OTHER THAN LIGHT SUPERFICIAL SCRATCHES THAT WOULD COME FROM NORMAL WEAR AND TEAR. FUNCTIONAL INSPECTION: PER THE ATTACHED PICTURE IN THIS SAP COMPLAINT, THREE DIFFERENT STYLE BLADES (RUSCH DISPOSABLE GREEN LITE MILL 2, SNAP LIGHT MILL 4, AND EMERALD RUSCH MIL 1.5) WERE ATTACHED TO EACH OF THE FIVE SUSPECTED DEFECTIVE HANDLES AND REMOVED WITHOUT ANY REASONABLE FORCE. ROOT CAUSE: NO FUNCTIONAL ISSUES FOUND.
THE COMPLAINT WAS REPORTED AS: THE CUSTOMER ALLEGES THAT THEY ARE UNABLE TO REMOVE THE BLADE FROM THE HANDLE WITHOUT EXTRAORDINARY EFFORT. THE HANDLES ARE USED WITH SUNMED, PHILLIPS AND RUSCH BLADES. THE END-USER STATES THAT THE ISSUE OCCURS OVER TIME. NO REPORT OF PT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 283977 | RUSCH FIBEROPTIC LARYNGOSCOPE HANDLE | FIBEROPTIC LARYNGOSCOPE HANDLE | CCW | TELEFLEX MEDICAL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | SUNMED BLADES| RUSCH BLADES| PHILLIPS BLADES |