FDA Adverse Event Malfunction Summary report: N

MEGA NEEDLE DRIVER INSTRUMENT

MDR report key: 3190548 · Received June 26, 2013

Report

Report Number
2955842-2013-02297
Event Type
Malfunction
Date Received
June 26, 2013
Date of Event
January 8, 2013
Report Date
May 31, 2013
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING EVALUATION FOUND THAT THE ONE GRIP CLOSE CABLE WAS BROKEN NEAR THE PROXIMAL PULLEYS AND PROXIMAL CLEVIS CABLE HOLE. THE IDLER PULLEY SPUN FREELY AND DID NOT EXHIBIT DAMAGE. THE CABLE SEGMENT STUCK OUT AT THE WRIST. NO MAJOR WEAR WAS FOUND ON THE PROXIMAL CLEVIS CABLE HOLE. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI SI PARTIAL NEPHRECTOMY PROCEDURE, THE MEGA NEEDLE DRIVER INSTRUMENT CABLE BROKE. NO MISSING OR FALLEN PIECES WERE REPORTED. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
291283 MEGA NEEDLE DRIVER INSTRUMENT ENDOSCOPIC INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420194-08 M10121004 557

Patients

Seq Age Sex Outcome Treatment
1 76 YR DA VINCI SI SYSTEM, INSTRUMENTS AND ACCESSORIES