FDA Adverse Event Malfunction Summary report: N

LARGE NEEDLE DRIVER INSTRUMENT

MDR report key: 3190543 · Received June 26, 2013

Report

Report Number
2955842-2013-02295
Event Type
Malfunction
Date Received
June 26, 2013
Date of Event
May 29, 2013
Report Date
May 31, 2013
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING EVALUATION FOUND THAT THE GRIP CABLE WAS FRAYED AT THE DISTAL IDLER PULLEY. THE FRAYED STRANDS STUCK OUT AT THE WRIST. THE OTHER CABLES AT THE WRIST WERE UNDAMAGED. THERE WERE INDENTATIONS AT THE EDGE OF THE DISTAL PULLEY. THERE WERE VISIBLE SCRATCHES ON THE SURFACE OF THE PULLEY. THE EVIDENCE WAS INCONCLUSIVE BUT THE DAMAGE WAS MOST LIKELY DUE TO MISHANDLING. THE INSTRUMENTS AND ACCESSORIES USER MANUAL SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS O HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. O DO NOT USE AN INSTRUMENT TO CLEAN DEBRIS FROM ANOTHER INSTRUMENT INTRAOPERATIVELY. THIS MAY RESULT IN DAMAGE TO THE INSTRUMENTS OR OTHER UNINTENDED CONSEQUENCES, SUCH AS DISCONNECTION OF THE INSTRUMENT TIP. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI SI DONOR HEPATECTOMY PROCEDURE THE LARGE NEEDLE DRIVER INSTRUMENT CABLE SNAPPED. NO MISSING OR FALLEN PIECES WERE REPORTED. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
291279 LARGE NEEDLE DRIVER INSTRUMENT ENDOSCOPIC INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420006-06 M14121217 379

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI SI SYSTEM, INSTRUMENTS AND ACCESSORIES