FDA Adverse Event Malfunction Summary report: N

BRAVO

MDR report key: 3190502 · Received June 21, 2013

Report

Report Number
9710107-2013-00155
Event Type
Malfunction
Date Received
June 21, 2013
Report Date
June 23, 2013
Manufacturer
GIVEN IMAGING LTD.
Product Code
FFT
PMA / PMN Number
K002028
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: ONE DELIVERY SYSTEM WAS RETURNED TO MEDTRONIC FOR EVALUATION. THE CAPSULE WAS NOT RETURNED FOR INVESTIGATION. THE DELIVERY SYSTEM WAS INVESTIGATED VISUALLY FOR EXTERNAL DAMAGE. THE DELIVERY SYSTEM WAS BENT SLIGHTLY AT THE END OF THE HANDLE, BUT THE PLUNGER WAS NOT BROKEN. THE PLUNGER ROTATED MORE THAN 1/8 OF A TURN. THE EMERGENCY RELEASE ON THE DELIVERY SYSTEM WAS NOT IMPLEMENTED. THE WIRE THAT HOLDS THE CAPSULE WAS COMPLETELY OFF AND THE FOAM GASKET WAS IN GOOD CONDITION. THE DELIVERY SYSTEM DID NOT HAVE ANY VISIBLE DAMAGE. AS THE PRODUCT WAS RECEIVED, THE PLUNGER ROTATED MORE THAN 1/8 OF A TURN WHICH IS INDICATIVE OF MISHANDLING. THERE WAS NO REQUIRED INTERVENTION TO PREVENT PERMANENT IMPAIRMENT/DAMAGE IN THIS CASE. THE INFORMATION WAS UPDATED IN THIS SUPPLEMENTAL.

Description of Event or Problem · 1

CUSTOMER REPORTED ON A BRAVO PH CAPSULE THAT FAILED TO ATTACH. THE PATIENT WAS NOT INJURED FOLLOWING THE PROCEDURE.

Description of Event or Problem · 1

A REPEAT PROCEDURE WAS NECESSARY DUE TO THE ALLEGED DEVICE MALFUNCTION. INTERVENTION WAS NOT REQUIRED. THERE WAS NOTHING UNUSUAL ABOUT THE PATIENT OR PROCEDURE ITSELF THAT MAY HAVE LED TO THIS EVENT. AN ENDOSCOPY WAS PERFORMED PRIOR TO THE PROCEDURE AND THE ESOPHAGUS APPEARED TO BE NORMAL. THE SITE HAS BEEN PERFORMING THIS PROCEDURE FOR ABOUT FIVE YEARS. NO KNOWN ADVERSE EVENTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
283819 BRAVO BRAVO PH CAPSULE FFT GIVEN IMAGING LTD. FGS-0313 21057Q

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention