FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 3190379 · Received June 26, 2013

Report

Report Number
1416980-2013-16442
Event Type
Malfunction
Date Received
June 26, 2013
Date of Event
June 1, 2013
Report Date
June 1, 2013
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THIS REPORT IS FOR SYSTEM ERROR 2240, WHERE THE HEATER BAG WAS NOT CONNECTED TIGHTLY ENOUGH TO THE HEATER LINE. USE ERRORS AND PROPER USER INSTRUCTIONS ARE ADDRESSED IN "THE HOMECHOICE AND HOMECHOICE PRO APD SYSTEMS PATIENT AT-HOME GUIDE", WHICH IS SHIPPED WITH EVERY HOMECHOICE DEVICE. IT INSTRUCTS THE USER TO CHECK ALL DISPOSABLE SET CONNECTIONS FOR A SECURE FIT BEFORE BEGINNING THERAPY. IT PROVIDES STEP-BY-STEP INSTRUCTIONS FOR CONNECTING THE SOLUTION BAGS. A REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. IF ANY ADDITIONAL RELEVANT INFORMATION IS OBTAINED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE HOME PATIENT (HP) HAD A SYSTEM ERROR 2240 (AIR IN TUBING) ON THE HOMECHOICE (HC). IT WAS UNKNOWN WHEN THE ALARM HAD OCCURRED. THE CAREGIVER (CG) STATED THAT THE HEATER BAG WAS NOT CONNECTED TIGHTLY ENOUGH TO THE HEATER LINE, AND A LEAK OCCURRED. THERE WAS NO REPORT OF PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE EVENT ASSOCIATED WITH THIS REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
290773 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 55 YR HOMECHOICE