NEUROFORM 3 EZ- 4.5 X 20MM
Report
- Report Number
- 3008853977-2013-00196
- Event Type
- Injury
- Date Received
- June 26, 2013
- Date of Event
- May 18, 2013
- Report Date
- May 31, 2013
- Manufacturer
- STRYKER NEUROVASCULAR CORK
- Product Code
- NJE
- PMA / PMN Number
- H020002
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
Narratives
THE SUBJECT DEVICE IS NOT AVAILABLE; THEREFORE ANALYSIS CANNOT BE PERFORMED. FROM THE INFORMATION AVAILABLE, THERE IS NO INDICATION THAT THE REPORTED EVENT IS RELATED TO PRODUCT SPECIFICATIONS NONCONFORMANCE OR MISUSE. THERE IS ALSO NO INDICATION THAT THE SUBJECT DEVICE MALFUNCTIONED. HOWEVER, PATIENT COMPLICATIONS IS A KNOWN RISK ASSOCIATED WITH SUCH PROCEDURES AND NOTED AS SUCH IN THE DIRECTIONS FOR USE (DFU). THEREFORE, A ROOT CAUSE OF ANTICIPATED PROCEDURAL COMPLICATION HAS BEEN ASSIGNED TO THE EVENT. THE PROBABLE CAUSE OF THE REPORTED STENT MIGRATION IS PROCEDURAL FACTORS AS IT IS LIKELY THAT PROCEDURAL FACTORS DURING USE LIMITED DEVICE PERFORMANCE.
(B)(4): SUBJECT DEVICE REMAINS IMPANTED.
IT WAS REPORTED THAT THE PATIENT UNDERWENT A PROCEDURE TO TREAT AN ANEURYSM LOCATED IN THE LEFT INTERNAL CAROTID ARTERY. THREE STENTS WERE USED TO COVER THE NECK OF THE ANEURYSM. WHEN THE PHYSICIAN APPROACHED THE ANEURYSM WITH THE MICROCATHETER TO DEPLOY A FOURTH STENT, THE CATHETER PUSHED THE PROXIMAL END OF THE THIRD STENT. THE PROXIMAL END OF THE THIRD STENT (THIS REPORT) MIGRATED INTO THE ANEURYSM. AS A RESULT, THE PHYSICIAN CHANGED THE TREATMENT STRATEGY FROM STENT ASSISTED TECHNIQUE TO USING COILS TO OCCLUDE THE INTERNAL CAROTID ARTERY. AFTER THE PROCEDURE, THE PATIENT EXPERIENCED RIGHT SIDE PARALYSIS AND APHASIA. NO FURTHER INFORMATION WAS PROVIDED.
IT WAS REPORTED THAT THE PATIENT UNDERWENT A PROCEDURE TO TREAT AN ANEURYSM LOCATED IN THE LEFT INTERNAL CAROTID ARTERY. THREE STENTS WERE USED TO COVER THE NECK OF THE ANEURYSM. WHEN THE PHYSICIAN APPROACHED THE ANEURYSM WITH THE MICROCATHETER TO DEPLOY A FOURTH STENT, THE CATHETER PUSHED THE PROXIMAL END OF THE THIRD STENT. THE PROXIMAL END OF THE THIRD STENT (THIS REPORT) MIGRATED INTO THE ANEURYSM. AS A RESULT, THE PHYSICIAN CHANGED THE TREATMENT STRATEGY FROM STENT ASSISTED TECHNIQUE TO USING COILS TO OCCLUDE THE INTERNAL CAROTID ARTERY. AFTER THE PROCEDURE, THE PATIENT EXPERIENCED RIGHT SIDE PARALYSIS AND APHASIA. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 289810 | NEUROFORM 3 EZ- 4.5 X 20MM | STENT, INTRAVASCULAR, NEUOVASCULAR | NJE | STRYKER NEUROVASCULAR CORK | 15393946 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention | 4 GDC COILS (STRYKER)| 2 NEUROFORM EZ STENTS (STRYKER)| 29 COILS (MANUFACTURER UNKNOWN) |