CD HORIZON SPINAL SYSTEM
Report
- Report Number
- 1030489-2013-02526
- Event Type
- Malfunction
- Date Received
- June 26, 2013
- Date of Event
- May 7, 2013
- Report Date
- May 28, 2013
- Manufacturer
- MDT SOFAMOR DANEK PUERTO RICO MFG
- Product Code
- KWP
- PMA / PMN Number
- K981676
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER
Narratives
(B)(6). (B)(4). NEITHER THE DEVICE NOR APPLICABLE IMAGING FILMS WERE RETURNED TO THE MANUFACTURER FOR EVALUATION, THEREFORE, THE CAUSE OF THE EVENT CANNOT BE DETERMINED.
ADDITIONAL INFO:: MACROSCOPIC AND OPTICAL EXAMINATION REVEALED THREAD CREST AND FLANK DAMAGE; THIS DAMAGE APPEARS TO HAVE INITIATED AT THE START OF THE THREAD, AND IS CONSISTENT AROUND THE DAMAGED PORTION OF THE THREAD. FUNCTIONAL EVALUATION WITH A SAMPLE MAS FOUND THE IMPLANT UNABLE TO BE FULLY ENGAGED INTO THE MAS HEAD. THE ABOVE OBSERVATIONS ARE CONSISTENT WITH MISALIGNMENT OF THE MAS AND SET SCREW THREADS DURING CONSTRUCT ASSEMBLY.
IT WAS REPORTED THAT HE PATIENT UNDERWENT A POSTERIOR SPINAL SURGICAL PROCEDURE. IT WAS REPORTED THAT THE SETSCREW SLIPPED DURING INSERTION INTO THE BONE SCREW. THE SETSCREW WAS REMOVED AND REPLACED. NO PATIENT COMPLICATIONS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 290817 | CD HORIZON SPINAL SYSTEM | APPLIANCE, FIXATION, SPINAL INTERLAMINAL | KWP | MDT SOFAMOR DANEK PUERTO RICO MFG | NA | H13A6806 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |