FDA Adverse Event Malfunction Summary report: N

LIGACLIP** ENDOSCOPIC APPLIER

MDR report key: 3190317 · Received June 26, 2013

Report

Report Number
3005075853-2013-03288
Event Type
Malfunction
Date Received
June 26, 2013
Date of Event
June 18, 2013
Report Date
June 20, 2013
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
FZP
PMA / PMN Number
K830503
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RECEIVED IN GOOD VISUAL CONDITION. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. UPON FUNCTIONAL TESTING OF THE DEVICE, THE INSTRUMENT LOADED, RETAINED AND DEPLOYED 6 CLIPS AS INTENDED. THE INSTRUMENT WAS FULLY FUNCTIONAL AND CONFORMING TO OUR MANUFACTURING REQUIREMENTS. NO CONCLUSION COULD BE REACHED AS TO WHAT MAY HAVE CAUSED THE REPORTED INCIDENT. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION WAS NOT PROVIDED BY THE INITIAL CONTACT. INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, FROM THE FIRST DAY, THE JAWS OF THE APPLIER WERE LOOSE WHICH WAS LEADING TO MALFORMATION OF CLIPS. UNKNOWN HOW CASE WAS COMPLETED. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
289615 LIGACLIP** ENDOSCOPIC APPLIER FZP ETHICON ENDO-SURGERY, LLC. NA A8527

Patients

Seq Age Sex Outcome Treatment
1