FDA Adverse Event
Other
Summary report: N
CLINITEK STATUS
MDR report key: 3190267
·
Received June 19, 2013
Report
- Report Number
- 1217157-2013-00104
- Event Type
- Other
- Date Received
- June 19, 2013
- Date of Event
- June 4, 2013
- Report Date
- June 4, 2013
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS INC.
- Product Code
- KQO
- PMA / PMN Number
- K031947
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- NURSE
Narratives
Additional Manufacturer Narrative · 1
CUSTOMER INDICATED THAT THE PATIENT WAS ON AZO. AS STATED IN THE IFU: SUBSTANCES THAT CAUSE ABNORMAL URINE COLOR MAY AFFECT THE READABILITY OF TEST PADS ON URINALYSIS REAGENT STRIPS. THESE SUBSTANCES INCLUDE VISIBLE LEVELS OF BLOOD OR BILIRUBIN AND DRUGS CONTAINING DYES (E.G., PYRIDIUM, AZO GANTRISIN, AZO GANTANOL), NITROFURANTOIN (MACRO DAN TIN, FURADANTIN), OR RIBOFLAVIN. CUSTOMER WAS INFORMED THAT ANY DISCOLORED SAMPLES SHOULD NOT BE RUN BY THIS METHOD BECAUSE IT IS A COLORMETRIC TEST.
Description of Event or Problem · 1
CUSTOMER REPORTS THAT A MULTISTIX 10SG DIPSTICK READ ERRONEOUSLY AS A MULTISTIX PRO 10LB DIPSTICK ON THE INSTRUMENT. THERE WAS NO REPORT OF INJURY FOR THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 278486 | CLINITEK STATUS | CT STATUS | KQO | SIEMENS HEALTHCARE DIAGNOSTICS INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|