FDA Adverse Event Injury Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3190230 · Received June 26, 2013

Report

Report Number
2531779-2013-09006
Event Type
Injury
Date Received
June 26, 2013
Report Date
May 29, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

LOSS OF PRIME ALARMS ARE NOT LIKELY TO CAUSE AN ADVERSE EVENT, AS THE PUMP WILL ALARM TO ALERT THE USER OF THE ISSUE. THE OWNER¿S BOOKLET INSTRUCTS THE USER TO BE PREPARED TO GIVE HIS OR HERSELF AN INJECTION OF INSULIN IF DELIVERY IS INTERRUPTED FOR ANY REASON. THE PUMP HAS NOT BEEN REQUESTED FOR RETURN. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

ON (B)(6) 2013 THE REPORTER CONTACTED ANIMAS ALLEGING THAT THE PUMP KEPT EMITTING LOSS OF PRIME ALARMS AND THE PATIENT SUBSEQUENTLY EXPERIENCED ELEVATED BLOOD GLUCOSE (BG) OVER 400MG/DL WITH NAUSEA AND VOMITING. THE PATIENT WAS REPORTEDLY GIVEN CORRECTION INJECTIONS TO BRING BG DOWN. DURING TROUBLESHOOTING, THE REPORTER NOTED A PATTERN OF THE LOSS OF PRIME ALARMS BEING AROUND TIMES OF INCREASED ACTIVITY. THE REPORTER NOTED THAT THE PATIENT SWINGS THE PUMP AND LIKES TO ¿TWIRL THE PUMP ROUND AND ROUND¿. CUSTOMER TECHNICAL SUPPORT ADVISED THAT THIS COULD LOOSEN THE CARTRIDGE CAP AND CAUSE LOSS OF PRIMES. THIS COMPLAINT IS BEING REPORTED BASED ON THE ALLEGATION THAT THE PATIENT EXPERIENCED HYPERGLYCEMIA WHILE USING INSULIN PUMP THERAPY DUE TO USE ERROR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
290716 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 10 YR Life Threatening