FDA Adverse Event Malfunction Summary report: N

SITTER SELECT

MDR report key: 3190192 · Received June 10, 2013

Report

Report Number
2020362-2013-00261
Event Type
Malfunction
Date Received
June 10, 2013
Report Date
May 7, 2013
Manufacturer
J. T. POSEY CO.
Product Code
KMI
PMA / PMN Number
EXEMPT
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

RESULTS: EVALUATION OF THE RETURNED UNIT DID NOT CONFIRM THE REPORTED ISSUE OF A CONTINUOUS SOUND WHEN IN USE WITH THE SENSOR. THE ALARM HAS NO POWER. HOWEVER, A BATTERY SPRING IS BADLY BENT AND THE UNIT POWERED ON WITH AN EXTERNAL POWER SUPPLY. THE UNIT PASSES ALL FUNCTIONAL TESTS. NOTE: INSTRUCTIONS FOR USE STATES: HOLD ALARM VERTICALLY UPSIDE DOWN TO PREVENT BATTERY SPRING FROM BENDING DURING BATTERY INSTALLATION. INSERT FOUR (4) NEW "AA" ALKALINE BATTERIES. TAKE CARE NOT TO DAMAGE BATTERY CONTACTS. (B)(4).

Description of Event or Problem · 1

CUSTOMER REPORTED THAT THE ALARM SOUNDS CONTINUOUSLY WHEN IN USE WITH THE PIR SENSOR. BATTERIES HAVE BEEN REPLACED WITH A NEW SUPPLY. NO VISIBLE DAMAGE TO THE OUTSIDE OF THE ALARM. CUSTOMER IS NOT SURE IF THE SENSOR IS FUNCTIONING PROPERLY AND DID NOT PROVIDE A DATE WHEN THIS WAS DISCOVERED. NO PT INCIDENT OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
262136 SITTER SELECT KMI J. T. POSEY CO. 8361 NA

Patients

Seq Age Sex Outcome Treatment
1 UNK PIR SENSOR PART #8351, LOT # E2008240030