SITTER SELECT
Report
- Report Number
- 2020362-2013-00261
- Event Type
- Malfunction
- Date Received
- June 10, 2013
- Report Date
- May 7, 2013
- Manufacturer
- J. T. POSEY CO.
- Product Code
- KMI
- PMA / PMN Number
- EXEMPT
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- OTHER
Narratives
RESULTS: EVALUATION OF THE RETURNED UNIT DID NOT CONFIRM THE REPORTED ISSUE OF A CONTINUOUS SOUND WHEN IN USE WITH THE SENSOR. THE ALARM HAS NO POWER. HOWEVER, A BATTERY SPRING IS BADLY BENT AND THE UNIT POWERED ON WITH AN EXTERNAL POWER SUPPLY. THE UNIT PASSES ALL FUNCTIONAL TESTS. NOTE: INSTRUCTIONS FOR USE STATES: HOLD ALARM VERTICALLY UPSIDE DOWN TO PREVENT BATTERY SPRING FROM BENDING DURING BATTERY INSTALLATION. INSERT FOUR (4) NEW "AA" ALKALINE BATTERIES. TAKE CARE NOT TO DAMAGE BATTERY CONTACTS. (B)(4).
CUSTOMER REPORTED THAT THE ALARM SOUNDS CONTINUOUSLY WHEN IN USE WITH THE PIR SENSOR. BATTERIES HAVE BEEN REPLACED WITH A NEW SUPPLY. NO VISIBLE DAMAGE TO THE OUTSIDE OF THE ALARM. CUSTOMER IS NOT SURE IF THE SENSOR IS FUNCTIONING PROPERLY AND DID NOT PROVIDE A DATE WHEN THIS WAS DISCOVERED. NO PT INCIDENT OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 262136 | SITTER SELECT | KMI | J. T. POSEY CO. | 8361 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | PIR SENSOR PART #8351, LOT # E2008240030 |