FDA Adverse Event Malfunction Summary report: N

TRANSTAR STRETCHER

MDR report key: 3190119 · Received June 11, 2013

Report

Report Number
1824206-2013-03030
Event Type
Malfunction
Date Received
June 11, 2013
Date of Event
May 13, 2013
Report Date
May 13, 2013
Manufacturer
HILL-ROM, INC.
Product Code
FPO
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN REPLACED ALL BRAKE CASTERS AND THE BRAKE HEX ROD TO RESOLVE THE ISSUE.

Description of Event or Problem · 1

THE TECHNICIAN ALLEGED THE BRAKE CASTERS ARE RATCHETING WHILE IN BRAKE MODE. NO PATIENT IMPACT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
264650 TRANSTAR STRETCHER WHEELED STRETCHER FPO HILL-ROM, INC. 8050D3844

Patients

Seq Age Sex Outcome Treatment
1