MAYFIELD INFINITY SKULL CLAMP
Report
- Report Number
- 3004608878-2013-00123
- Event Type
- Injury
- Date Received
- June 24, 2013
- Date of Event
- June 17, 2013
- Report Date
- June 24, 2013
- Manufacturer
- INTEGRA LIFESCIENCES CORPORATION OH/USA
- Product Code
- HBL
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
TO DATE, THE DEVICE (MAYFIELD INFINITY SKULL CLAMP ALONG WITH THE A 1083 SKULL PINS) INVOLVED IN THE REPORTED INCIDENT HAS NOT BEEN REC'D FOR EVALUATION. AN INVESTIGATION HAS BEEN INITIATED BASED UPON THE REPORTED INFO.
IT WAS REPORTED THAT THE ROCKER ARM SLIPPED, CAUSING A DEEP LACERATION. ADD'L INFO WAS REQUESTED AND THE FOLLOWING WAS PROVIDED ON (B)(6) 2013: THE MODEL NUMBER OF THE SKULL CLAMP WAS PRODUCT ID A1114 (INFINITY MAYFIELD HEAD CLAMP). THE (B)(6) FEMALE PT WAS TO UNDERGO A LAMINECTOMY FOR REMOVAL OF A SACRAL TUMOR. THE PT WAS ON A (B)(4) TABLE WITH CERVICAL MANAGEMENT SYSTEM AND THE INFINITY MAYFIELD HEAD CLAMP. THE SURGERY WAS NOT PERFORMED WITH A STEREOTAXY DEVICE. INTEGRA DISPOSABLE ADULT SKULL PINS (A1083) WERE USED. LOT NUMBER OF THE SKULL PINS WAS NOT AVAILABLE. THE PT WAS IN A PRONE POSITION FOR SURGERY AND WAS NOT REPOSITIONED AT ANY TIME DURING THE SURGERY. IT WAS NOTED THAT THERE WAS BLEEDING COMING FROM THE PT'S HEAD AND AT THAT TIME, THE PT WAS TURNED SUPINE ONTO THE STRETCHER. THERE WAS NO NOTABLE SHIFTING OF THE HEADHOLDER NOTED BY THE NEUROSURGICAL RESIDENT. THE LENGTH OF TIME THE PRODUCT IN USE BEFORE THE EVENT OCCURRED WAS 15 MINUTES. THE PT SUSTAINED THE INJURY PRIOR TO THE START OF THE PROCEDURE. THE LACERATION WAS LEFT ANTERIOR, 5 MM IN LENGTH DISCOVERED AFTER PT HAD BEEN TURNED PRONE ONTO THE JACKSON TABLE. THE PT'S HEAD WAS SECURED INTO THE (B)(4) CERVICAL MANAGEMENT SYSTEM AND PT WAS BEING PADDED WITH FOAM IN PREPARATION FOR SURGERY. THE LACERATION WAS CLOSED WITH NON-INTEGRA VICRYL SUTURE AND THERE WAS NO VISIBLE OR REPORTED EFFECTS BY THE PT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 285330 | MAYFIELD INFINITY SKULL CLAMP | SKULL CLAMPS AND HEADREST SYSTEMS | HBL | INTEGRA LIFESCIENCES CORPORATION OH/USA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 50 YR | Required Intervention |