FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 3190087 · Received June 18, 2013

Report

Report Number
1720753-2013-07322
Event Type
Malfunction
Date Received
June 18, 2013
Date of Event
June 12, 2013
Report Date
June 19, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE REPRESENTATIVE EVALUATED THE SYSTEM AND REPLACED THE CINE DRIVE, INTERCONNECT CABLE, AND LEMO RECEPTACLE. THE SYSTEM OPERATES AS INTENDED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE MAINFRAME HAD "X-RAYS DISABLED ERROR MESSAGE." THE SYSTEM WOULD BOOT UP BUT IMMEDIATELY DISPLAY THE ERRORS AND THE CUSTOMER COULD NOT BYPASS THE ERRORS. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THE EVENT DESCRIBED IN THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
276011 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1