FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 3190078 · Received June 18, 2013

Report

Report Number
1720753-2013-07326
Event Type
Malfunction
Date Received
June 18, 2013
Date of Event
June 13, 2013
Report Date
June 19, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE REPRESENTATIVE EVALUATED THE SYSTEM AND ADJUSTED THE 5 VOLT POWER SUPPLY, AND REPLACED THE GENERATOR INTERFACE BOARD. THE SYSTEM OPERATES AS INTENDED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE SYSTEM DISPLAYED A COMMUNICATION ERROR. THIS ERROR WILL CAUSE THE SYSTEM TO LOCKUP, SHUT DOWN, OR NOT TO BOOT. NO PATIENT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
275969 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1