FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 3190063 · Received June 18, 2013

Report

Report Number
1720753-2013-07336
Event Type
Malfunction
Date Received
June 18, 2013
Date of Event
June 10, 2013
Report Date
June 19, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE POWER SUPPLY VOLTAGE WAS ADJUSTED. THE SYSTEM WAS THEN FOUND TO BE OPERATING AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE LEFT MONITOR WOULD NOT TURN ON AND THEY COULD NOT SEE THE LIVE IMAGE. THERE IS NO REPORT OF PATIENT INJURY ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
276246 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1