FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 3190055 · Received June 18, 2013

Report

Report Number
1720753-2013-07343
Event Type
Malfunction
Date Received
June 18, 2013
Date of Event
June 13, 2013
Report Date
June 19, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ON SITE INVESTIGATION. THE FLUORO FUNCTIONS BOARD WAS REPLACED. THE SYSTEM WAS THEN FOUND TO BE OPERATING AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM FAILED DURING A CASE AND DISPLAYED AN X-RAY DISABLED ERROR MESSAGE. THIS ERROR MESSAGE PREVENTS THE SYSTEM FROM PRODUCING X-RAY. THERE IS NO REPORT OF PT INJURY ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
276659 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1