FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 3190048 · Received June 18, 2013

Report

Report Number
1720753-2013-07352
Event Type
Malfunction
Date Received
June 18, 2013
Date of Event
June 16, 2013
Report Date
June 19, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE POWER SUPPLY PLUGS WERE RESEATED AND ALL GENERATOR BOARDS WERE RESEATED. THE SYSTEM WAS THEN FOUND TO BE OPERATING AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM WAS UNABLE TO PRODUCE FLUOROSCOPY AND WAS DISPLAYING AN X-RAY DISABLED ERROR MESSAGE. THERE IS NO REPORT OF PATIENT INJURY ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
276107 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1