FDA Adverse Event Malfunction Summary report: N

7700

MDR report key: 3190001 · Received June 18, 2013

Report

Report Number
9680959-2013-01203
Event Type
Malfunction
Date Received
June 18, 2013
Date of Event
March 22, 2013
Report Date
June 19, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS GMBH
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE GENERATOR BOARD WAS REPLACED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE SYSTEM WOULD NOT MAKE X-RAYS. THIS RESULTED A TOTAL LOSS OF IMAGING FUNCTIONALITY. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
276639 7700 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS GMBH 7700

Patients

Seq Age Sex Outcome Treatment
1