2517506-2013-00275
Report
- Report Number
- 2517506-2013-00275
- Event Type
- Malfunction
- Date Received
- June 26, 2013
- Date of Event
- May 28, 2013
- Report Date
- May 29, 2013
- PMA / PMN Number
- K070853
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- MEDICAL TECHNOLOGIST
Narratives
ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE FALSELY ELEVATED ETOH RESULTS IS UNKNOWN. TESTING ON THE OPEN WELL STABILITY OF A RETAINED SAMPLE FROM THE SAME LOT OF FLEX(R) REAGENT CARTRIDGES AS USED BY THE CUSTOMER WAS PERFORMED IN THE SIEMENS HEALTHCARE DIAGNOSTICS INC. TECHNICAL SERVICE LABORATORY TO CONFIRM THE FIVE DAY CLAIM IN THE REAGENT INSTRUCTIONS FOR USE. NO INCREASE IN ETOH RECOVERY WAS OBSERVED OVER FIVE DAYS WITH OPEN WELL CONDITIONS. THE COMPLAINT WAS NOT CONFIRMED. THE IFU FOR THE ETHYL ALCOHOL FLEX REAGENT CARTRIDGES LIMITATIONS OF PROCEDURE SECTION CONTAINS THE FOLLOWING INFORMATION: ENVIRONMENTAL ETHANOL CONTAMINATION MAY RESULT IN ELEVATED ETHANOL RESULTS AND HAS BEEN SHOWN TO EFFECT OPEN WELL STABILITY.
A FALSELY ELEVATED ETHYL ALCOHOL (ETOH) RESULT WAS OBTAINED ON A PATIENT'S SAMPLE. THE RESULT WAS REPORTED TO THE PHYSICIAN WHO QUESTIONED THE RESULT. THE PATIENT SAMPLE WAS REPEATED WITH A NEW REAGENT FROM A PREVIOUSLY UNOPENED FLEX(R) WELL COMPARTMENT AND A NEGATIVE RESULT WAS OBTAINED. IT IS UNKNOWN IF PATIENT TREATMENT WAS ALTERED OR PRESCRIBED ON THE BASIS OF THE ELEVATED ETOH RESULT. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE FALSELY ELEVATED ETOH RESULT.
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |