Narratives
CURRENT INFORMATION IS INSUFFICIENT TO PERMIT A CONCLUSION AS TO THE CAUSE OF THE EVENT. REVIEW OF DEVICE HISTORY RECORDS SHOW THAT LOT RELEASED WITH NO RECORDED ANOMALY OR DEVIATION. THERE ARE WARNINGS IN THE PACKAGE INSERT THAT STATE THAT THIS TYPE OF EVENT CAN OCCUR: UNDER POSSIBLE ADVERSE EFFECTS, NUMBER 15 STATES, "ELEVATED METAL ION LEVELS HAVE BEEN REPORTED WITH METAL ON METAL ARTICULATING SURFACES." THIS REPORT IS NUMBER 3 OF 4 MDRS FILED FOR THIS EVENT (REFERENCE 1825034-2013-02115, 02115-1, 02228 AND 02229 ).
LEGAL COUNSEL FOR THE PATIENT REPORTED THAT PATIENT UNDERWENT LEFT TOTAL HIP ARTHROPLASTY ON (B)(6) 2008. LEGAL COUNSEL FOR PATIENT REPORTS PATIENT ALLEGATIONS OF PERSONAL INJURY. THERE HAS BEEN NO REPORTED REVISION PROCEDURE. LABORATORY REPORT PROVIDED INDICATES ELEVATED CHROMIUM LEVEL. THIS REPORT IS BASED ON ALLEGATIONS SET FORTH IN PLAINTIFF¿S COMPLAINT AND THE ALLEGATIONS CONTAINED THEREIN ARE UNVERIFIED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 291196 | M2A-MAGNUM 42-50MM TPR INSRT-3 | PROSTHESIS, HIP | KWA | BIOMET ORTHOPEDICS | N/A | 536600 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization |