FDA Adverse Event Other Summary report: N

PILLCAM SB2 CAPSULE

MDR report key: 3187976 · Received June 21, 2013

Report

Report Number
9710107-2013-00137
Event Type
Other
Date Received
June 21, 2013
Report Date
June 18, 2013
Manufacturer
GIVEN IMAGING LTD.
Product Code
NEZ
PMA / PMN Number
K101250
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

GIVEN IMAGING LABELING, SUCH AS USER MANUAL INDICATE THAT PILLCAM SB CAPSULES ARE CONTRA-INDICATED FOR USE IN SOME PATIENTS. ALTHOUGH THE PATIENTS HAD NO KNOWN HISTORY OF SWALLOWING PROBLEMS AND THE CAPSULE INGESTION WAS UNEVENTFUL, ONE MAY SUSPECT, HOWEVER, THAT THE PATIENT DOES INDEED HAVE A SWALLOWING PROBLEM IF HE DID ASPIRATE THE CAPSULE. SWALLOWING DISORDER IS A RECOGNIZED CONTRA-INDICATION TO THE USE OF CAPSULE ENDOSCOPY.

Description of Event or Problem · 1

A PATIENT UNDERWENT PILLCAM CAPSULE ENDOSCOPY PER NORMAL PROTOCOL WITHOUT ANY INCIDENT. THE PHYSICIAN WHO READ THE PILLCAM STUDY STATED THAT IT WAS AN INCOMPLETE STUDY AND THE CAPSULE DID NOT LEAVE THE STOMACH. FOUR MONTH LATER FOLLOWING PATIENT'S COMPLAINTS OF SHORTNESS OF BREATH, IT WAS REVEALED IN AN X-RAY THAT THE CAPSULE APPEARED TO BE IN THE PATIENT'S LUNGS. THE CAPSULE WAS RETRIEVED BY BRONCHOSCOPY AND THE PATIENT IS WELL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
282317 PILLCAM SB2 CAPSULE PILLCAM SB CAPSULE NEZ GIVEN IMAGING LTD. PILLCAM SB2 CAPSULE NOT AVAILABLE

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention