FDA Adverse Event Malfunction Summary report: N

HOMECHOICE PRO

MDR report key: 3183708 · Received June 22, 2013

Report

Report Number
1416980-2013-16225
Event Type
Malfunction
Date Received
June 22, 2013
Date of Event
June 2, 2013
Report Date
June 2, 2013
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WV, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). A REVIEW OF THE DEVICE LOGS FOUND THAT DRAIN VOLUMES MET INTRA-PERITONEAL VOLUME (IIPV) CRITERIA. DURING DEVICE EVALUATION, THE DEVICE PASSED HOMECHOICE RITE ELECTRICAL AND RITE FUNCTIONAL TESTING. THE EXTERNAL AND INTERNAL INSPECTIONS FOUND NO ISSUES. THE DEVICE HAD SUCCESSFULLY PASSED MULTIPLE SIMULATED THERAPIES. THE ROOT CAUSE OF THE REPORTED ISSUE WAS DETERMINED TO BE INSUFFICIENT DRAIN DUE TO ONE OR MORE CYCLES ADVANCES TO THE NEXT FILL WHEN SLOW OR NO FLOW OCCURRED ABOVE THE MINIMUM DRAIN VOLUME THRESHOLD. NO FAILURE OR MALFUNCTION WAS IDENTIFIED WITH THE DEVICE THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED ISSUE. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A HOME PATIENT (HP) RECEIVED A HIGH DRAIN 105 ALARM ON THE HOMECHOICE (HC) MACHINE DURING USE, WHILE THE HP WAS NOT CONNECTED. THE HIGH DRAIN ALARM INDICATES THAT THE PATIENT DRAINED GREATER THAN 200% OF THE MAXIMUM PRESCRIBED CYCLE FILL VOLUME IN STANDARD MODE, MEETING INCREASED INTRA-PERITONEAL VOLUME (IIPV) CRITERIA. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) ADVISED THE HP TO USE MANUAL SUPPLIES IN ORDER TO FINISH THE THERAPY. THERE WAS NO REPORT OF PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE EVENT ASSOCIATED WITH THIS REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
284225 HOMECHOICE PRO SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1 43 YR HOMECHOICE AUTOMATED PD SET WITH CASSETTE| DIABETES-RELATED MEDICATION