MATRIX SCREW Ø1.85 SELF-TAP L5 TAN 1U I/
Report
- Report Number
- 2520274-2013-03403
- Event Type
- Malfunction
- Date Received
- June 22, 2013
- Date of Event
- February 1, 2012
- Report Date
- February 3, 2012
- Manufacturer
- SYNTHES USA
- Product Code
- JEY
- PMA / PMN Number
- K083388
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- DA
- Reporter Occupation
- OTHER
Narratives
SYNTHES IS SUBMITTING THIS REPORT AS A RESULT OF REMEDIATION ACTIVITIES RELATED TO FDA WARNING LETTER DATED FEBRUARY 2012. DEVICE(S) LISTED IN THIS REPORT IS (ARE) USED FOR TREATMENT, NOT DIAGNOSIS. ANY ADDITIONAL INFORMATION RECEIVED REGARDING THIS EVENT AFTER FILING THIS REPORT SHALL BE FILED ON A SUPPLEMENTAL MDR. A REVIEW OF THE DEVICE HISTORY RECORDS WAS NOT PERFORMED BECAUSE NO LOT NUMBER WAS PROVIDED AND IT WAS NOT AVAILABLE ON THE RETURNED PRODUCT. THE ADDITIONAL EVALUATION REVEALED THAT THE SCREW HAS CLEARLY VISIBLE STRESS MARKS AND THAT THE THREAD FLANK IS PARTIALLY SHEARED OFF. THE MEASURABLE DIMENSIONS OF THE SCREW AND THE PLATES WERE AS FAR AS POSSIBLE CHECKED AND FOUND ACCORDING TO THE SPECIFICATION. OUR INVESTIGATION HAS SHOWN THAT SUCH AN OCCURRENCE CAN EITHER HAPPEN BY AN EXCESSIVE CONTACT WITH THE PLATE WHEN THE SCREW IS INSERTED IN AN EXTREME ANGLE OR WHEN THE PLATE WAS DAMAGED BY DRILLING PREVIOUSLY. IN THE ENCLOSED PICTURES IT IS VISIBLE THAT THE PLATE IS DAMAGED. IT IS IMPOSSIBLE TO DETERMINE IF THE DAMAGES WERE CAUSED BY AN EXCESSIVE SCREW CONTACT OR THE DRILL BIT.
IT WAS REPORTED THAT DURING AN ORTHOGNATHIC SURGERY USING MATRIX ORTHOGNATHIC PLATE AND SCREW, THE SURGEON DISCOVERED A METAL THREAD WHEN INSERTING THE SCREW. THE SURGEON TURNED THE SCREW BACK AND THE THREAD APPEARED. IT WAS IN THE POSTERIOR PART OF THE LEFT SIDE OF THE MANDIBLE. THE HOLE IN THE SAGITTAL PLATE IS MARKED WITH AN ELASTIC BAND. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 284315 | MATRIX SCREW Ø1.85 SELF-TAP L5 TAN 1U I/ | JEY | SYNTHES USA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 31 YR |