FDA Adverse Event Malfunction Summary report: N

OT PING METER

MDR report key: 3183408 · Received June 22, 2013

Report

Report Number
3008382007-2013-18037
Event Type
Malfunction
Date Received
June 22, 2013
Report Date
June 16, 2013
Manufacturer
LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
Product Code
NBW
PMA / PMN Number
K082590
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1 (08/13/2013)-PRODUCT EVALUATION: THE ANALYSIS OF THE SUBJECT METER WAS COMPLETED ON (B)(4) 2013 BY PRODUCT ANALYSIS WITH THE FOLLOWING FINDINGS: THE ANALYSIS OF THE SUBJECT METER FOUND A RAM FAILURE WITHIN THE DEVICE. THE ANALYSIS FOUND THAT THE METER DISPLAYS PC WHEN POWER WAS ON. RAM IC105 WAS FOUND TO BE DEFECTIVE. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.

Additional Manufacturer Narrative · 1

LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT.

Description of Event or Problem · 1

ON (B)(6) 2013, THE REPORTER CONTACTED LIFESCAN USA, ALLEGING AN UNUSUAL ISSUE WHERE THE METER WOULD DISPLAY A MESSAGE SAYING "PC". THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. REPLACEMENT PRODUCTS WERE SENT TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
284335 OT PING METER GLUCOSE MONITORING SYS/KIT NBW LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL 3444187

Patients

Seq Age Sex Outcome Treatment
1 20 YR