FDA Adverse Event Malfunction Summary report: N

UNICEL® DXC 600I SYNCHRON® ACCESS® CLINICAL SYSTEM

MDR report key: 3183328 · Received June 21, 2013

Report

Report Number
2050012-2013-00426
Event Type
Malfunction
Date Received
June 21, 2013
Date of Event
May 26, 2013
Report Date
May 26, 2013
Manufacturer
BECKMAN COULTER
Product Code
JJE
PMA / PMN Number
K060256
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
MEDICAL TECHNOLOGIST

Narratives

Additional Manufacturer Narrative · 1

UPON FURTHER TROUBLESHOOTING, THE CUSTOMER REVEALED A LOOSE REAGENT PROBE B FITTING ON TOP OF THE LEAKING PROBE. THE CUSTOMER RESOLVED THE ISSUE BY TIGHTENING THE LOOSE FITTING FOUND ON TOP OF REAGENT PROBE B. THE CUSTOMER ALSO REPLACED THE AST REAGENT CARTRIDGE. SERVICE WAS NOT DISPATCHED FOR THIS EVENT AS THE CUSTOMER RESOLVED THE ISSUE. (B)(4).

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER (BEC) REPORTING THAT THE UNICEL DXC 600I SYNCHRON ACCESS CLINICAL INSTRUMENT WAS GENERATING "CHEMISTRY CARTRIDGE (CC) CUVETTE NOT DRY BEFORE 1ST REAGENT DISPENSE" ERRORS. UPON TROUBLESHOOTING WITH BEC HOTLINE SUPPORT, THE CUSTOMER OBSERVED THAT REAGENT PROBE B WAS LEAKING INTO THE DRIP TRAY. IN ADDITION, THE CUSTOMER DISCOVERED LIQUID ON TOP OF THE ASPARTATE AMINOTRANSFERASE (AST) REAGENT CARTRIDGE WHICH ATTRIBUTED TO THE LEAKING REAGENT PROBE. THE LEAK WAS CONTAINED WITHIN THE INSTRUMENT. THE CUSTOMER WAS WEARING PERSONAL PROTECTIVE EQUIPMENT (PPE) CONSISTING OF GLOVES, A LABORATORY COAT, AND PROTECTIVE EYEWEAR AT THE TIME OF THE INCIDENT. THE MATERIAL SAFETY DATA SHEET (MSDS) WAS NOT REVIEWED. THERE IS AN EXPOSURE CONTROL / RISK MANAGEMENT PLAN IN PLACE AT THE FACILITY. THERE WAS NO BIOHAZARD EXPOSURE TO MUCOUS MEMBRANES OR OPEN WOUNDS. MEDICAL ATTENTION WAS NOT SOUGHT. NO ERRONEOUS RESULTS WERE GENERATED IN CONNECTION TO THIS EVENT.