FDA Adverse Event Injury Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3183270 · Received June 21, 2013

Report

Report Number
3004209178-2013-94921
Event Type
Injury
Date Received
June 21, 2013
Date of Event
June 1, 2013
Report Date
June 2, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NE
Reporter Occupation
PATIENT FAMILY MEMBER OR FRIEND

Narratives

Additional Manufacturer Narrative · 1

A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING INCLUDING THE DISPLACEMENT TEST. THE DEVICE WAS RECEIVED WITH INTERMITTENT BUTTON RESPONSE DUE TO MOISTURE DAMAGE KEYPAD TRACES.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER WAS HOSPITALIZED DUE TO DIABETES KETOACIDOSIS. THE FATHER STATED THAT THE BUTTONS KEYPADS WERE STICKING, AND HE REQUESTED A REPLACEMENT. THE CUSTOMER'S BLOOD GLUCOSE READING AT TIME OF HER ADMISSION WAS 260MG/DL. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
284193 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-722NAL

Patients

Seq Age Sex Outcome Treatment
1 19 YR Hospitalization