PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 3004209178-2013-94923
- Event Type
- Injury
- Date Received
- June 21, 2013
- Date of Event
- June 1, 2013
- Report Date
- June 2, 2013
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY
- Reporter Occupation
- PATIENT FAMILY MEMBER OR FRIEND
Narratives
A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING INCLUDING THE DISPLACEMENT TEST. SEVERAL BOLUSES WERE PROGRAMMED AND DELIVERED. THE BOLUSES WERE RECORDED PROPERLY IN THE BOLUS HISTORY SCREEN. AFTER TESTING IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS. THE DEVICE HAD SCRATCHED DISPLAY WINDOW.
IT WAS REPORTED THAT THE CUSTOMER WAS IN THE EMERGENCY ROOM DUE TO HIGH BLOOD GLUCOSE OF 460MG/DL, AND HER BLOOD GLUCOSE WAS TREATED WITH AN INSULIN DRIP. IT WAS STATED THAT THE CUSTOMER HAD MISSING BOLUS, WHICH CAUSED HER GLUCOSE LEVEL TO RISE. THE FATHER STATED THAT THE CUSTOMER BOLUSED, BUT IT WAS NOT RECORDED IN THE BOLUS HISTORY. THE CALLER DID NOT FEEL COMFORTABLE AND REQUESTED A REPLACEMENT. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 282868 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-723NAB |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 11 YR | Hospitalization |