FDA Adverse Event Malfunction Summary report: N

LEGEND FOOTED ATTACHMENT

MDR report key: 3183192 · Received June 21, 2013

Report

Report Number
1625507-2013-00031
Event Type
Malfunction
Date Received
June 21, 2013
Report Date
May 23, 2013
Manufacturer
MDT POWERED SURGICAL SOLUTIONS
Product Code
HBB
PMA / PMN Number
K020069
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

REPORT CONFIRMED. EVALUATION DETERMINED THAT THE FOOTED PORTION WAS DAMAGED BY TOOL CONTACT. A PORTION OF THE FOOT OF THE ATTACHMENT WAS DETACHED AND MISSING. THE LIKELY CAUSES ARE IDENTIFIED AS DEBRIS IN THE COLLET AND IMPROPER INSERTION OF THE TOOL. (B)(4) WAS INITIATED TO INVESTIGATE THIS MALFUNCTION. THE USER MANUAL CONTAINS THE FOLLOWING WARNING ¿THE ATTACHMENT SHOULD NOT BE USED IF ANY PART OF THE ATTACHMENT APPEARS TO BE BENT, LOOSE, MISSING, OR DAMAGED. EXCESSIVE PRESSURE OR IMPROPER HANDLING, SUCH AS BENDING OR PRYING, OF THE ATTACHMENT OR DISSECTING TOOL MAY CAUSE INJURY TO THE PATIENT, OPERATOR AND/OR OPERATING ROOM STAFF." IN ADDITION, ¿A TACTILE AND AUDIBLE CLICK WILL BE OBSERVED WHEN ATTACHMENT IS FULLY SEATED. TACTILE FEEDBACK IS FELT WHEN THE DISSECTING TOOL IS FULLY SEATED. TURN THE TOOL LOCKING RING TO THE ¿LOCK¿ POSITION. VERIFY THAT THE TOOL IS IN PLACE BY GENTLY PULLING ON THE TOOL.¿ THE PREVENTIVE MAINTENANCE/SERVICE MANUAL FOR THE LEGEND SYSTEM SPECIFIES SERVICE INTERVALS FOR DEVICES BASED ON THE HOSPITAL U SAGE LEVEL. THE MAXIMUM SPECIFIED SERVICE INTERVAL IS 24 MONTHS. DEVICE HAS BEEN IN USE FOR APPROXIMATELY 35 MONTHS WITH NO RECORD OF FACTORY SERVICE DURING THIS PERIOD. WE WILL CONTINUE TO MONITOR THIS COMPLAINT TYPE FOR TRENDS. (B)(4).

Description of Event or Problem · 1

DEVICE RETURNED FOR NON-SPECIFIED REPAIR. NO PATIENT IMPACT REPORTED. REPAIR ESCALATED TO COMPLAINT ON EVALUATION DUE TO ATTACHMENT DAMAGED. NO ADDITIONAL INFORMATION WAS OBTAINED ON FOLLOW - UP.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
283346 LEGEND FOOTED ATTACHMENT MOTOR, DRILL, PNEUMATIC HBB MDT POWERED SURGICAL SOLUTIONS N/A N/A

Patients

Seq Age Sex Outcome Treatment
1