FDA Adverse Event Injury Summary report: N

XIENCE V EVEROLIMUS ELUTING CORONARY STENT SYSTEM

MDR report key: 3182895 · Received June 21, 2013

Report

Report Number
2024168-2013-03872
Event Type
Injury
Date Received
June 21, 2013
Date of Event
May 29, 2013
Report Date
May 29, 2013
Manufacturer
AV-TEMECULA-CT
Product Code
NIQ
PMA / PMN Number
P070015
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. IT IS INDICATED THAT THE DEVICE IS NOT RETURNING FOR EVALUATION; THEREFORE A FAILURE ANALYSIS OF THE COMPLAINT DEVICE COULD NOT BE COMPLETED. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. REVIEW OF THE COMPLAINT HISTORY OF THE REPORTED LOT DID NOT INDICATE A MANUFACTURING ISSUE. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A MILDLY TORTUOUS, MILDLY CALCIFIED, AND 95% STENOSED MID LEFT ANTERIOR DESCENDING ARTERY. PRE-DILATATION WAS PERFORMED WITH A TREK BALLOON CATHETER. WHILE ADVANCING A 2.75 X 23 MM XIENCE V PARTIALLY THROUGH THE LESION IT GOT STUCK AND THERE WAS STRONG RESISTANCE PULLING BACK THE CATHETER, SO THE STENT WAS IMPLANTED PARTIALLY IN THE LESION. PLAIN OLD BALLOON ANGIOPLASTY (POBA) WAS PERFORMED USING AN NC TREK TO TREAT THE REST OF THE LESION. THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
283986 XIENCE V EVEROLIMUS ELUTING CORONARY STENT SYSTEM DRUG ELUTING CORONARY STENT SYSTEM NIQ AV-TEMECULA-CT 2051741

Patients

Seq Age Sex Outcome Treatment
1 62 YR Required Intervention