FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 3182880 · Received June 21, 2013

Report

Report Number
1416980-2013-16172
Event Type
Malfunction
Date Received
June 21, 2013
Date of Event
May 30, 2013
Report Date
May 31, 2013
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS NOT AVAILABLE FOR EVALUATION. THE HOME PATIENT HAD DISCONNECTED FROM THE SUPPLY LINE AND CONNECTED TO THE PATIENT LINE, WHICH IS A USE ERROR THAT CAN CAUSE AIR TO ENTER THE PERITONEAL CAVITY AND/OR CAUSE CONTAMINATION. USE ERRORS AND PROPER USER INSTRUCTIONS ARE ADDRESSED IN THE HOMECHOICE AND HOMECHOICE PRO APD SYSTEMS PATIENT AT-HOME GUIDE, WHICH IS SHIPPED WITH EVERY HOMECHOICE DEVICE. THIS GUIDE PROVIDES STEP-BY-STEP INSTRUCTIONS FOR SETTING UP THE HOMECHOICE FOR THERAPY AND GIVES A VISUALIZATION OF THE SUPPLY LINES AND PATIENT LINE. THE GUIDE PROVIDES STEP-BY-STEP INSTRUCTIONS FOR PROPERLY CONNECTING THE PATIENT LINE TO THE TRANSFER SET. A FORMAL REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. IF ADDITIONAL RELEVANT INFORMATION BECOMES AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING A CALL FOR THERAPY ASSISTANCE, IT WAS FOUND THAT THE HOME PATIENT (HP) HAD CONNECTED HIMSELF TO THE SUPPLY LINE, STARTED THERAPY, AND THEN DISCONNECTED FROM THE SUPPLY LINE AND CONNECTED TO THE PATIENT LINE. THE HP WAS IN INITIAL DRAIN AT THE TIME OF THE CALL. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) ASSISTED THE CAREGIVER (CG) TO END THERAPY EARLY AND ADVISED THE CG TO NOTIFY THEIR NURSE AND START OVER WITH NEW SUPPLIES. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS INDICATED IN ASSOCIATION WITH THIS REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
283963 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 HOMECHOICE