PARADIGM REAL-TIME INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2013-02516
- Event Type
- Malfunction
- Date Received
- June 21, 2013
- Date of Event
- May 30, 2013
- Report Date
- May 30, 2013
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY
- Reporter Occupation
- PATIENT
Narratives
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
THE INSULIN PUMP UNABLE TO PRIME DURING PRIME TEST DUE TO FAULTY FORCE SENSOR. UNABLE TO PERFORM THE OCCLUSION TEST DUE TO PRIME ANOMALY.
CUSTOMER STATED THAT THERE IS A PROBLEM WITH THE SUPPLIES OR INSULIN PUMP. CUSTOMER'S BLOOD GLUCOSE READING IS 301 MG/DL. CUSTOMER HAS TREATED WITH BOLUS. DURING TROUBLESHOOTING, TIME AND DATE ARE CORRECT. PROGRAMMING IS CORRECT. MANUAL PRIME IS CORRECT, THE INSULIN DID EXIT THE TUBING. HIGH PRESSURE TEST FAILED TWICE. THE INSULIN PUMP WILL BE REPLACED. NOTHING FURTHER REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 283879 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MMT-522RNAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 78 YR |