FDA Adverse Event
Malfunction
Summary report: N
6000CMS
MDR report key: 3182641
·
Received June 11, 2013
Report
- Report Number
- 1722139-2013-01771
- Event Type
- Malfunction
- Date Received
- June 11, 2013
- Date of Event
- January 15, 2011
- Report Date
- July 13, 2011
- Manufacturer
- MOOG DEVICE GROUP
- Product Code
- FRN
- PMA / PMN Number
- K981816
- Removal / Correction Number
- Z-1867-2011
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
INVESTIGATION FOUND THAT PUMP HAD EXPERIENCED ERROR CODE 36 IN PUMP'S HISTORY. NEW PUMP SOFTWARE WAS INSTALLED. REFERENCE RECALL NUMBER Z-1867-2011.
Description of Event or Problem · 1
INFORMATION RECEIVED STATES THAT PUMP HAD EXPERIENCED ERROR CODE 36.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 264608 | 6000CMS | FRN | MOOG DEVICE GROUP | 6000CMS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |