FDA Adverse Event Malfunction Summary report: N

COBAS 6000 CORE

MDR report key: 3182632 · Received June 21, 2013

Report

Report Number
1823260-2013-03765
Event Type
Malfunction
Date Received
June 21, 2013
Date of Event
June 12, 2013
Report Date
July 25, 2013
Manufacturer
ROCHE DIAGNOSTICS
Product Code
JGS
PMA / PMN Number
K060373
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

IT WAS UNKNOWN IF THE INITIAL REPORTER SENT REPORT TO THE FDA. RESULTS -DEVICE SUBASSEMBLY-VACUUM TUBING.

Additional Manufacturer Narrative · 1

THE INVESTIGATION COULD NOT DETERMINE A SPECIFIC ROOT CAUSE DUE TO THE LACK OF RELEVANT INFORMATION. THE CUSTOMER COULD NOT PROVIDE RESULTS FOR CHLORIDE AND POTASSIUM FOR THE SAMPLES. IT WAS NOTED THAT THE RESULTS MEASURED ON THE C501 ANALYZER CANNOT SIMPLY BE COMPARED TO THE RESULTS FROM I-STAT. THE C501 MEASURES ISES INDIRECTLY, WHILE THE I-STAT MEASURES ISE DIRECTLY. IT WAS ALSO NOTED THAT THE CHANGE IN THE VACUUM TUBING WOULD BE REASONABLE IF OTHER TESTS WERE LOW.

Description of Event or Problem · 1

THE CUSTOMER RECEIVED QUESTIONABLE RESULTS FOR ION SELECTIVE ELECTRODE (ISE) SODIUM (NA) ON FIVE PATIENTS. OF THOSE FIVE, IT WAS DETERMINED THAT ONE PATIENT HAD ERRONEOUS NA RESULTS THAT WERE REPORTED OUTSIDE OF THE LABORATORY. ALL RESULTS ARE IN MMOL/L. THE PATIENT'S INITIAL NA RESULT WAS 125, WHICH WAS REPORTED OUTSIDE OF THE LABORATORY. THE SAMPLE WAS REPEATED ON AN ISTAT ANALYZER AND GENERATED A REPEAT NA RESULT OF 131. THE CUSTOMER DEEMED THE REPEAT RESULT TO BE THE CORRECT RESULT AND ISSUED A CORRECTED REPORT. THERE WAS NO ADVERSE EVENT. THE LOT NUMBER OF THE NA ELECTRODE IN USE WAS NOT PROVIDED BY THE CUSTOMER. THE FIELD SERVICE REPRESENTATIVE FOUND A HOLE IN THE RINSE NOZZLE VACUUM TUBING. HE REPLACED THE TUBING. THE CUSTOMER PERFORMED CALIBRATION AND QC AND ALL RESULTS WERE ACCEPTABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
282768 COBAS 6000 CORE CLINICAL CHEMISTRY ANALYZER JGS ROCHE DIAGNOSTICS NA NA

Patients

Seq Age Sex Outcome Treatment
1