FDA Adverse Event Malfunction Summary report: N

COAGUCHEK ® XS SYSTEM

MDR report key: 3182629 · Received June 21, 2013

Report

Report Number
1823260-2013-03761
Event Type
Malfunction
Date Received
June 21, 2013
Date of Event
April 29, 2013
Report Date
June 27, 2013
Manufacturer
ROCHE DIAGNOSTICS
Product Code
GJS
PMA / PMN Number
K062925
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

IT WAS UNKNOWN IF THE INITIAL REPORTER SENT REPORT TO THE FDA.

Description of Event or Problem · 1

CALLER REPORTS THAT DURING A CORRELATION STUDY THE FOLLOWING COAGUCHEK XS/LABORATORY RESULTS WERE OBTAINED: 7.9 INR/5.01 INR, 3.8 INR/2.88 INR, 6.4 INR/2.68 INR, 4.8 INR/3.34 INR, 6.6 INR/3.43 INR, 4.0 INR/2.84 INR, 4.8 INR/3.42 INR. NO ACTIONS TAKEN BASED ON DEVICE RESULTS. NO ADVERSE EVENT REPORTED. REQUESTED RETURN OF SUSPECT SYSTEM AND REPLACEMENT WAS SENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
282767 COAGUCHEK ® XS SYSTEM PROTHROMBIN TIME TEST STRIPS GJS ROCHE DIAGNOSTICS NA 21611011

Patients

Seq Age Sex Outcome Treatment
1