AJUST ADJUSTABLE SINGLE INCISION SLING
Report
- Report Number
- 1018233-2013-02431
- Event Type
- Malfunction
- Date Received
- June 14, 2013
- Date of Event
- May 15, 2013
- Report Date
- May 15, 2013
- Manufacturer
- BARD SHANNON LIMITED
- Product Code
- FTL
- PMA / PMN Number
- K092607
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
THE REPORTED EVENT COULD NOT BE CONFIRMED. NO SAMPLE WAS RETURNED FOR EVAL. BASED ON THE INVESTIGATION FINDINGS, CURRENT MFG CONTROLS ARE CONSIDERED ADEQUATE AS TO DETECT AND SEGREGATE ANY NON-CONFORMING UNIT. EVENT DESCRIBED COULD NOT BE CONFIRMED AS A MFG RELATED ISSUE. THE INSTRUCTIONS FOR USE WHICH ACCOMPANIES ALL DEVICES CURRENTLY ADDRESSES POTENTIAL RISKS ASSOCIATED WITH SURGICALLY IMPLANTED MATERIALS. THE INSTRUCTIONS FOR USE STATES IN THE ADVERSE EVENTS: "COMPLICATIONS ASSOCIATED WITH THE PROPER IMPLANTATION OF THE AJUST SLING SYSTEM MAY INCLUDE, BUT ARE NOT LIMITED TO: POSTOPERATIVE HEMATOMA, SEROMA, ABSCESS OR FISTULA FORMATION, OR SCARRING WHICH MAY OCCUR FOLLOWING THE IMPLANT PROCEDURE. URINARY RETENTION, BLADDER OUTLET OBSTRUCTION AND OTHER VOIDING DYSFUNCTIONS. THESE CONDITIONS MAY BE ASSOCIATED WITH OVER-CORRECTION/TOO MUCH TENSION PLACED ON THE IMPLANT. PERFORATIONS OR LACERATIONS OF VESSELS, NERVES, BLADDER, BOWEL, URETHRA, OR ANY VISCERA, WHICH MAY OCCUR DURING THE IMPLANTATION PROCEDURE. IRRITATION AT THE OPERATIVE WOUND SITE WHICH MAY ELICIT A FOREIGN BODY RESPONSE THAT LEADS TO WOUND DEHISCENCE, INFLAMMATION AND/OR INFECTION. EXTRUSION THROUGH VAGINAL EPITHELIUM OR EROSION INTO SURROUNDING VISCERA AND/OR MUCOSA. INFLAMMATION, SENSITIZATION, PAIN, DYSPAREUNIA, SCARIFICATION, CONTRACTION, DEVICE MIGRATION AND FAILURE OF THE PROCEDURE RESULTING IN RECURRENCE OF INCONTINENCE." (B)(4).
(B)(4). IT WAS REPORTED THAT THE FIXED ANCHOR BROKE DURING THE PLACEMENT OF THE SLING. HOWEVER, THE DOCTOR WAS STILL ABLE TO IMPLANT THE SLING SUCCESSFULLY WITHOUT ANY FURTHER IMPACT TO THE PT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 272390 | AJUST ADJUSTABLE SINGLE INCISION SLING | FTL | BARD SHANNON LIMITED | NA | HUWH1795 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |