FDA Adverse Event
Malfunction
Summary report: N
9800
MDR report key: 3182495
·
Received June 14, 2013
Report
- Report Number
- 1720753-2013-07141
- Event Type
- Malfunction
- Date Received
- June 14, 2013
- Date of Event
- June 3, 2013
- Report Date
- June 14, 2013
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SVC REP PERFORMED AN ON SITE INVESTIGATION. THE BOARDS WERE RESEATED AND THE LEMO CONNECTOR WAS REPLACED DURING THE SVC CALL. THE SYS WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SVC.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE SYS HAD A COMMUNICATION FAILURE ERROR MESSAGE. THIS ERROR MAY RESULT IN A SYS LOCK UP, NO BOOT, OR SHUT DOWN SITUATION. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THE EVENT DESCRIBED IN THIS COMPLAINT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 272095 | 9800 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9800 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |