FDA Adverse Event
Malfunction
Summary report: N
9800
MDR report key: 3182491
·
Received June 14, 2013
Report
- Report Number
- 1720753-2013-07143
- Event Type
- Malfunction
- Date Received
- June 14, 2013
- Date of Event
- June 4, 2013
- Report Date
- June 14, 2013
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SVC REP PERFORMED AN ON SITE INVESTIGATION. THE REPORTED ISSUE COULD NOT BE DUPLICATED. THE INTERCONNECT CABLE WAS REPLACED DURING THE SVC CALL. THE SYS WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SVC.
Description of Event or Problem · 1
IT WAS REPORTED THE SYS HAD INTERMITTENT COMMUNICATIONS ERRORS AND WOULD NOT BOOT UP. THERE IS NO REPORT OF PT INJURY OR DEATH.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 272243 | 9800 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9800 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |