FDA Adverse Event
Malfunction
Summary report: N
7900
MDR report key: 3182403
·
Received June 14, 2013
Report
- Report Number
- 9680959-2013-01166
- Event Type
- Malfunction
- Date Received
- June 14, 2013
- Date of Event
- June 3, 2013
- Report Date
- June 14, 2013
- Manufacturer
- GE OEC MEDICAL SYSTEMS GMBH
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE REPRESENTATIVE EVALUATED THE SYSTEM AND RESEATED CIRCUIT BOARDS, CABLES, AND CONNECTORS. THE SYSTEM OPERATES AS INTENDED.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT INTERMITTENTLY, THERE WAS A GENERATOR ERROR. THIS MESSAGE APPEARS WHEN THE SYSTEM DETECTS AN ERROR IN ANY OF THE REGISTERS ASSOCIATED WITH THE HIGH VOLTAGE GENERATOR. THE SYSTEM AUTOMATICALLY SHUTS DOWN WHEN THIS OCCURS. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THE EVENT DESCRIBED IN THIS COMPLAINT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 271897 | 7900 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS GMBH | 7900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|